Vaccum Drying Oven (Four Shelf)

Vaccum Drying Oven (Four Shelf)
Request a Quote
Main technical parameter index(can be measured at an ambient temperature of +5 to +35 °C)
Structure
  • Working chamber size: 590mm height, 660mm width, 1275mm depth. 4 chambers.
  • Dimensions: Refer to the schematic diagram.
  • Insulation thickness: 90mm.
  • Supporting foot: 100mm.
  • Heater position: Around the inside of the cavity.
  • Pumping port position: 4 chambers are concentrated in one suction port and vacuumed, controlled by solenoid valve
  • Venting port position: Equipped with inflatable solenoid valve.
  • Opening method: 4 single doors.
  • Control box position: Right side.
  • Silicone Rubber high temperature resistance sealing ring.
  • Equipped with water cooling cycle system.
  • Each chamber is equipped with a stainless steel trolley.
  • Each chamber has a blast wheel on the back, magnetic fluid connection.
Main Property
  • Working temperature: Room temperature +15 ° C ~ 200 ° C
  • Instrument accuracy: ±1℃
  • Temperature fluctuation: ±1°C
  • Heating time: ≤120min(empty box,25℃~150℃)
  • Temperature control method: PID adjustment
  • Vacuum degree: ≥100Pa (0.0999MPa)
  • Average air leakage in 24 hours: ≤0.3kPa/h;
  • Outside surface temperature: ≤ room temperature +25 ° C
Electrical capacity:
  • Power supply: AC380V 3-phase 4-wire 50Hz
  • Heating power: 8KW*4
  • Blast power: 370W*4
  • Maximum current: 20A*4
  • Air switch capacity:100A
Security devices:
  • Air switch (prevent overcurrent and short circuit)
  • Overcurrent protection (prevent high current)
  • Over temperature protection (prevent high temperature, protect the product), over temperature alarm.
Functional requirements
  • The Q series substrate is program controlled, and display settings are shown on an 8 inch color LCD touch screen.
  • Equipped with 4 electromagnetic valves, the vacuuming, holding pressure, and inflating processes are fully automatic and program controlled.
  • Users can freely adjust the vacuum degree and temperature, and automatically maintain both under a certain temperature and vacuum degree range.
  • Equipment operation processes (vacuum, holding pressure, and inflation) are automatically PLC controlled.
Related products
Send Message
Inquiry
Inquiry